Title: Arterial blood gas, laboratory and
portable point-of-care testing for perioperative haemoglobin measurement in
patients undergoing major non-cardiac surgeries: A comparative study
Introduction:
Worldwide,
close to 85 million red blood cell (RBC) units are transfused annually, with
surgical patients accounting for up to 44% of transfusions.1–3 Clinically
indicated RBC transfusions in surgery can be life- saving interventions, while
inappropriate transfusions provide no clinical benefit and can cause harm.(1) Perioperative transfusions are guided by
blood loss assessment and hemoglobin (Hb) values. Hb measured by arterial blood
gas (ABG) analyser is convenient, easily available and results are obtained
faster than the conventional laboratory methods (Lab). (2) It is well established that hemoglobin measurement
plays a central role in any decision to transfuse RBCs. A recent survey of
Canadian anaesthesiologists reported intraoperative hemoglobin levels to be the
most important parameter for transfusion decision making- more important blood
loss or hemodynamics.(3) It is common practice to perform daily blood
analysis in intensive care units (ICUs), the results are delayed by a few hours
when processed by a central laboratory (Lab) . Blood gas analyzers have been
widely into modern ICUs and offer a unique opportunity to determine
measurements at the point-of-care (POC), and in a time frame of 2 minutes.
(4) Current practice guidelines suggest that the
decision to transfuse RBCs in surgery should be generally limited to those with
a haemoglobin (Hb) of 60–100 g/L and informed by important clinical parameters.
Implicit in this recommendation is the emphasis on measuring Hb at critical
times during surgery. Traditionally, Hb
measurement during surgery was done through submission of a complete blood
count specimen to the main laboratory. More recently, point- of- care testing
(POCT) instruments capable of measuring or calculating haemoglobin (POCT- Hgb)
have become commonplace in surgery due to their ease of use and rapid delivery
of a result. Few clinical trials have compared intraoperative transfusion
strategies, the majority of which have limited sample sizes. There are three
main classes of POCT-Hgb methods currently in use in operating rooms. The first
class yields a Hgb value from 10 µL of whole blood in less than 60 secs. Second
class can yield results using 65 µL to 100 µL of whole blood in 120 s using
conductometric methods and third class involves non-invasive monitoring of
capillary Hgb through multiwavelength sensors and pulse co-ox imetry.(1) As per the quality requirements based on
2024 CLIA (Clinical Laboratory Improvement Amendment) Acceptance Limits for
Proficiency Testing (Hematology CLIA 2024), the new criteria for AP for
Hemoglobin & Hematocrit is (Test Value) TV ± 4% for both.
To
date, POCT- Hgb instruments have been validated with static and normal Hgb
values, such as in healthy blood donors. In clinical care, they have been
examined primarily in the non- operative setting, such as the outpatient clinic
or emergency department, reporting conflicting results in relation to bias and
accuracy of POCT-Hgb devices. Their value in surgery, where Hgb can change
rapidly due to bleeding and hemodilution from concurrent intravenous fluid
administration, are relatively untested. More importantly, there is also a
paucity of published validation data based on intraoperative Hgb values within
the transfusion zone of 60–100 g/L used for clinical decision- making,
highlighting a major gap in existing evaluations.(1)
In
our study we plan to determine the reliability and accuracy of Hgb measurements
by ABG analyser and POCT-portable Hgb analyser compared to the laboratory
method during major non-cardiac surgeries and having a significant amount of
blood loss.
Research Question: Is ABG analyser & Point of care portable Hgb analyser accurate
in assessing Hgb levels in comparison to Central laboratory method?
Objectives:
Primary Objective:
To
compare the accuracy of Hgb measurements obtained by ABG analyser, POCT
Portable Hgb analyser with central laboratory method during major non-cardiac surgeries
Secondary Objective:
To
evaluate the level of agreement between the measurement techniques of ABG, POCT
portable Hgb analyser and laboratory method
Hypothesis:
The
hemoglobin values obtained by ABG analyser and POCT portable Hgb analyser might
show significant differences compared to the values obtained by the laboratory
method in patients having significant amount of intraoperative bleeding.
Methods:
Study design:
Non-randomised
Interventional
prospective comparative study
Study Setting:
Department
of Anaesthesiology, Critical care & Pain medicine, AIIMS Guwahati.
Study Population:
Inclusion criteria:
·
Age:
18-60 years
·
Undergoing
Major non-cardiac surgery with anticipated significant blood loss*
·
Preoperative
Hb>10g/dl
·
Patients
belonged to ASA Physical status Class I to III
*Operational
definition for significant
intraoperative blood loss for the present study is minimum of 70% of maximum
allowable blood loss.
Exclusion criteria:
·
Patients
with known hematologic disorders
·
patients
with severe coagulopathies
·
Uncontrolled
Hypertension
·
Patients
on diuretics and anticoagulants
·
Ischaemic
Heart disease
Sample Size:
Based
on Sunil Ranjan et al(2) using mean differences and standard deviation and
Marianne Johnson et al (5) where comparison of Hgb measurements by 3 Point of
care devices with standard laboratory values sample size for our study was
calculated.
In
order to achieve a power of 95%, with an alpha error of 0.05, and a 95%
confidence interval, with an effect size of 0.29, we will need approximately 548 samples. We added a 10% drop-out to
make the sample number to 603.75(Rounded off to 604). Given that each
participant provides 4 samples, 151
participants will be required to get 684 samples. The sample size was
calculated for cross-sectional study with Z alpha adjusted value of 2.41 and Z
beta of 1.96 for 95% power comparing
three techniques of Hgb estimation was 137. The sample size was calculated using open
epidemiological tool www.openepi.com.
In our
study, the Bonferroni correction will be applied to the p-values obtained from
the pairwise comparisons between the three techniques at each time point. If
there are multiple time points or additional comparisons, the total number of
comparisons will be considered.
Example Calculation
- Number of Techniques: 3
- Number of Comparisons: 3 (ABG
vs. POCT, ABG vs. Lab, POCT vs. Lab)
- Initial Alpha Level: 0.05
Applying
the Bonferroni correction:
Corrected alpha:
0.05/3=0.017
This
comprehensive approach ensures that the study is sufficiently powered and
accounts for potential dropouts, enhancing the reliability and validity of the
results.
After
applying the Bonferroni correction, any p-value less than 0.017 would be
considered statistically significant instead of the usual 0.05 threshold. This
correction helps to maintain the overall Type I error rate at the desired level
(5% in this case) across all comparisons. This will help ensure that the
conclusions drawn are statistically valid and not due to chance.
In our
study, We are enrolling 151 participants will be collecting 4 samples per participant.
This
comprehensive approach ensures that the study is sufficiently powered and
accounts for potential dropouts, enhancing the reliability and validity of the
results.
Procedure:
After satisfying
the inclusion criteria all patients are kept fasting for 6 hr preoperatively. After
shifting the patient to the operation theatre, standard monitors such as pulse
oximeter, non-invasive blood pressure monitor (NIBP), Temperature(oC)
and Electrocardiograph(ECG) are attached. Arterial cannulation under local
anaesthesia will be performed using 20G cannula or BD arterial cannula and two
large bore intravenous cannula will also be placed.(2) Arterial transducers will be used for
invasive hemodynamic monitoring in all non-cardiac surgeries anticipating
significant blood loss intraoperatively. Each time an intraoperative Hgb
measurement is under-taken by an anaesthesiologist, the patient will undergo
three concurrent Hgb measurements using the same sample.(5) The anaesthesiologist will draw a standard
3 ml of blood from the arterial line and run on the first device i.e POCT-portable
Hgb analyser (AGAPPE, MispaHbX portable hemoglobin analyzer) which assess
Hb/Hct by photometric detection method and about 0.5 to 1 ml of blood is
collected in a heparinised syringe and sent for ABG analysis on ABG analyzer (ESCHWEILER
Combi line 2, ESCHWEILER GmbH & Co.KG, Kiel, Germany) and remaining 2 ml of
blood is sent to the central laboratory in an EDTA vacutainer and tested using Hematology
analyzers used in our Institute laboratory i.e (SYSMEX XP-300 & SYSMEX
XN-1000) which uses non-cyanide method and Sulfolyser method for hemoglobin estimation
respectively. These samples are collected at four time points i.e
preoperatively before induction of anaesthesia, Intraoperatively at a point of
about 70% MABL(Maximal allowable blood loss), Postoperatively immediately in
the PACU/ICU and at 6 hrs in the PACU/ICU.
All
patients will receive anaesthesia following a standard protocol. After
induction patients will be hydrated with Ringer’s Lactate/Kabilyte at the maintenance
rate based on the weight of the patient using the 4-2-1 Estimation of maintenance
requirements based on data by Holliday & Segar. During episodes of excess
loss, such as when body cavities are open or bleeding occurs, the fluid volumes
prescribed are then based on perceived knowledge of the movement of fluids between
compartments
Intraoperative
blood loss will be assessed by weighing blood soaked surgical swabs, measuring
volume of blood loss in suction apparatus, and loss at surgical site will also
be taken into account. Up to MABL, intravascular volume will be replaced with
either Ringer’s lactate or Kabilyte (based on lactate levels) 3:1 ratio
and colloids like Hydroxy ethyl starch or albumin(if existing hypoalbuminemia) in 1:1
ratio as appropriate to replace blood loss. The Hgb levels <8 g/dl is used
as trigger point for transfusion of blood products when there is significant
intraoperative blood loss. Intraoperative use of vasopressors and packed red
blood cells/Fresh frozen plasma/Platelets transfusion will be documented
accordingly. Intraoperative hemodynamic parameters like HR/min, MAP mm/Hg will
be documented.(2) The time of sample collection T1, T2, T3
& T4 are documented and also the time of report collection is also
documented. The demographic data of the participant i.e Age in years , Sex,
Weight in Kg, Height in cm, BMI(Kg/m2, Type of Surgery,
Emergency/Elective, ASA class,
Preoperative Hgb, Arterial or Venous sample and Type of Anaesthesia are
documented. The number of IV fluids used intraoperatively (Crystalloids &
Colloids), Use of blood preventative methods (e.g., tranexaemic acid) and
amount of blood loss are thoroughly documented. The time of start of surgery
and time of end of surgery and also the duration of surgery are documented.
Missing Data:
Missing
data is expected with all the three methods of testing Hgb. In very rare
instances, there may be missing data for the lab-Hgb value due to clotted blood
sample or an insufficient blood sample. As the Lab-Hgb value is the reference
standard to assess the precision of POCT-Hgb device & ABG analyser, any Hgb
measurement missing data will be excluded from the analysis. Furthermore,
during the data collection phase of the study, blood samples with missing
lab-Hgb values will not be considered in the total sample size.
Missing
data for the POCT portable Hgb analyzer (Mispa HbX, AGAPPE) should be rare, but
might occur in instances of device failure (e.g.,if microcuvette is filled
inappropriately) or unavailability of device being used in another operating
room, or due to insufficient quantity to run all the tests. Considering the
very rare instance of missing data in this device, they will be excluded in the
analysis.
Reference test(Gold standard) test:
The
central lab Hgb sample will be collected in an EDTA vacuum collection tube and
delivered to the laboratory for processing in the usual fashion. The POCT
portable Hgb analyzer(AGAPPE, MispaHbX portable Hemoglobin Analyzer) and
Arterial Blood Gas analyzer(ESCHWEILER Combi-line 2) are considered Index
tests. For each device, the manufacturer’s instructions and institutional
procedures will be strictly adhered to, including device handling, training ,
storage, calibration and quality control.(5,6)
Blood gas analysis:
To
determine the Hgb level of either arterial or venous heparinised blood samples,
ESCHWEILER Combiline 2 Analyzer will be used in our Institute. Hemoglobin level
was quantified with using the law of Lambert-Beersch in a secondary analysis of
RCT study by Tanner L et al(7) in which If Hb was <0.16 g/dl (0.1 mmol/l) or >40.26
g/dl (25 mmol/l), results were classified as outlier and not considered in the
analysis. Blood samples, both arterial and venous, were taken by trained
medical staff and transferred to the ABG analyzer immediately. Results were
displayed in the patients medical file & documented.
Standardization
Techniques
- Calibration
and Quality Control:
- All
devices (ABG Analyzer, POCT Hemoglobin Analyzer, Central Laboratory
Analyzers) will undergo regular calibration and quality control as per
the manufacturer’s and institutional guidelines.
- Blood
Sample Handling:
- Samples
will be drawn and processed uniformly to minimize preanalytical
variability.
- Protocol
Adherence:
- Strict
adherence to study protocols for sample collection, handling, and
analysis to ensure consistency.
- Training:
- Personnel
involved in sample collection and analysis will receive standardized
training to ensure uniformity in procedures.
Outcomes:
Primary outcomes:
- To compare the accuracy of
hemoglobin (Hgb) measurements obtained by the ABG analyzer, POCT portable
Hgb analyzer with central laboratory method during major non-cardiac
surgeries
Secondary outcomes:
1.
Level of Agreement:
o
Assessment of the agreement between the Hgb measurement techniques
(ABG, POCT, and central lab) using statistical methods such as Bland-Altman
plots or Intraclass Correlation Coefficient (ICC).
Data Collection:
Data
will be recorded and documented in standardised forms by blinded outcome
assessor
Statistical Analysis:
- Descriptive Statistics:
- Mean,
median, standard deviation, range for Hgb values from each method.
- Paired t-test or Wilcoxon
Signed-Rank Test:
- To compare
the Hgb values between two measurement techniques (e.g., ABG vs. POCT,
ABG vs. laboratory, POCT vs. laboratory).
- Use the
paired t-test if data are normally distributed; otherwise, will use
Wilcoxon Signed-Rank test.
- ANOVA or Friedman Test:
- To compare
Hgb values across all three measurement techniques simultaneously.
- Use ANOVA
if the data are normally distributed; otherwise, will use the Friedman
test for non-parametric data.
Secondary
Objective
1.
To evaluate the level of agreement between the measurement techniques of ABG,
POCT portable Hgb analyzer with central laboratory method
- Bland-Altman
Plots was used to assess agreement between two measurement techniques.
- Intraclass
Correlation Coefficient (ICC)was used to assess the reliability and agreement
among the three measurement techniques.
Summary
of Statistical Tests
- Descriptive Statistics:
- Mean,
median, standard deviation, range
- Comparative Tests:
- Paired
t-test or Wilcoxon Signed-Rank Test
- ANOVA or
Friedman Test
- Agreement Tests:
- Bland-Altman
Plots
- Intraclass
Correlation Coefficient (ICC)
|